Variability of immunohistochemical reactivity on stored paraffin slides.

J Clin Pathol 1998 May;51(5):370-4

Laboratory of Pathology, Hospital Saint-Louis, Paris, France.

Aims: To investigate the effects of slide storage on immunohistochemical staining, since recent reports have indicated that storage of unstained paraffin slides for up to 12 weeks may lead to false negative immunostaining of tumour markers.

Methods: 11 antibodies (anti-cytokeratin, epithelial membrane antigen (EMA), vimentin, smooth muscle actin, PS100, chromogranin, CD45, CD20, CD3, CD30, and oestrogen receptor (OR) were tested on unstained paraffin slides of breast carcinomas, lymphomas, and neuroendocrine tumours that had been stored for three to 10 years. All the paraffin blocks were recut less than one week before immunostaining. Immunostainings of years old slides were compared with those of recent slides in at least five cases for each antibody. For three antibodies (antichromogranin, anti-CD3, and anti-OR) we also tested one year old and three months old slides.

Results: Intensity of staining on years old slides was strikingly reduced for chromogranin and CD3 in several cases and was slightly stronger for vimentin. In some cases a significant decrease of OR positivity was observed after three months storage, and a complete loss of OR immunostaining after 12 months. No significant difference was noted with the other antibodies.

Conclusions: Immunohistochemical detection of some antigens located either in the nucleus, in the cytoplasm, or on the cytoplasmic membrane could be impaired by storage of paraffin slides as short a time as three months. One should be cautious of doing retrospective immunohistochemical studies on stored unstained slides.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC500697PMC
http://dx.doi.org/10.1136/jcp.51.5.370DOI Listing
May 1998
2 Reads

Publication Analysis

Top Keywords

paraffin slides
16
three months
12
slides
8
years slides
8
unstained paraffin
8
three
5
paraffin
5
antibody three
4
cases antibody
4
anti-cd3 anti-or
4
antichromogranin anti-cd3
4
antibodies antichromogranin
4
three antibodies
4
anti-or tested
4
year three
4
intensity staining
4
staining years
4
slides strikingly
4
slidesresults intensity
4
months slidesresults
4

References

(Supplied by CrossRef)
Immunocytochemical methods and protocols. Preparation of frozen sections for analysis
Bratthauer, G.L. et al.
Methods in molecular biology 1994
p53 Antigen loss in stored paraYn slides
Prioleau, J.E.; Schnitt, S.J. et al.
N Engl J Med 1995
Loss of tumour marker–immunostaining intensity on stored paraYn slides of breast cancer
Jacobs, T.W.; Prioleau, J.E.; Stillman, I.E. et al.
J Natl Cancer Inst 1996
More on p53 antigen loss in stored paraYn slides
Kato, J.; Sakamaki, S.; Niitsu, Y. et al.
N Engl J Med 1995
Guardians of the wax... and the patient [editorial]
Mills, S.E.; Fechner, R.E.; Frierson, H.F. et al.
Am J Clin Pathol 1995
Loss of p53 immunostaining intensity in breast cancer
Henson, D.E. et al.
J Natl Cancer Inst 1996

Similar Publications