Patient-tailored SyncAV algorithm: A novel strategy to improve synchrony and acute hemodynamic response in heart failure patients treated by cardiac resynchronization therapy.

J Cardiovasc Electrophysiol 2020 02 25;31(2):512-520. Epub 2019 Dec 25.

Department of Cardiology, Shanghai Institute of Cardiovascular Diseases, Zhongshan Hospital, Fudan University, Shanghai, China.

Introduction: Several automatic algorithms have developed to optimize the timing cycle setting in cardiac resynchronization therapy (CRT). The present study aims to investigate whether the novel device-based SyncAV algorithm could elicit better synchrony and acute hemodynamic response.

Methods And Results: Thirty five patients undergoing CRT implantation were prospectively studied. The device was programmed to three biventricular (BiV) pacing modes sequentially after the procedure: QuickOpt algorithm (mode I), SyncAV algorithm with default 50 ms offset (mode II), and SyncAV algorithm with optimized offset minimizing QRS duration (QRSd) (mode III). After each setting, electrocardiographic and echocardiographic data were collected. As a result, QRSd was reduced from 172.8 ± 17.9 ms during intrinsic conduction to 153.1 ± 15.9 ms in mode I, further narrowed to 140.5 ± 16.7 ms in mode II, and reached shortest (134.8 ± 16.1 ms) in mode III (P < .01 for all). Besides, significantly shorter QT intervals were observed in mode I (453.2 ± 45.5 ms), mode II (443.9 ± 34.2 ms) and mode III (444.1 ± 28.7 ms), compared with native condition (472.5 ± 51.2 ms) (P < .01). All three BiV modes exhibited comparable T T interval and T T /QT ratio (P > .05). Mode I presented significantly higher aortic velocity time integral than intrinsic conduction (21.0 ± 6.4 cm vs 18.4 ± 5.5 cm; P < .01), which was even higher in mode II (22.0 ± 6.5 cm) and mode III (23.7 ± 6.5 cm). All three BiV modes significantly reduced standard deviation of time to peak contraction of 12-LV segments (Ts-SD) (Mode I: 55.2 ± 16.5 ms, Mode II: 50.2 ± 14.7 ms, Mode III: 45.4 ± 14.4 ms) compared with intrinsic conduction (66.3 ± 18.4 ms) (P < .01), with Mode III demonstrating the smallest (P < .01).

Conclusion: SyncAV CRT ameliorated electrical and mechanical synchrony as well as acute hemodynamic response beyond conventional QuickOpt optimization. An additional individualized adjustment to the SyncAV offset added to its advantage.

Download full-text PDF

Source
http://dx.doi.org/10.1111/jce.14315DOI Listing
February 2020

Publication Analysis

Top Keywords

syncav algorithm
16
resynchronization therapy
8
cardiac resynchronization
8
mode iii
8
acute hemodynamic
8
synchrony acute
8
mode syncav
8
mode
6
algorithm
5
setting electrocardiographic
4
algorithm elicit
4
elicit better
4
crt study
4
hemodynamic responsemethods
4
1728 ± 179 ms intrinsic
4
intrinsic conduction
4
better synchrony
4
device-based syncav
4
aims investigate
4
conduction 1531 ± 159 ms
4

Similar Publications