Calculating dose from a 2.5 MV imaging beam using a commercial treatment planning system.

J Appl Clin Med Phys 2019 Dec 1;20(12):25-35. Epub 2019 Nov 1.

Department of Medical Physics, School of Medicine and Public Health, University of Wisconsin-Madison, Madison, WI, 53705, USA.

Patient dose from 2.5 MV images on the TrueBeam linear accelerator is not easily quantified, primarily because this beam energy is not normally modeled by commercial treatment planning systems. In this work we present the feasibility of using the Eclipse® treatment planning system to model this beam. The Acuros XB and the AAA dose calculation algorithms were tested. Profiles, PDDs, and output factors were measured for the 2.5 MV unflattened imaging beam and used for beam modeling. The algorithms were subsequently verified using MPPG 5.a guidelines. Calculated doses with both algorithms agreed with the measurement data to within the following criteria recommended for conventional therapeutic MV beams: 2% local dose-difference in the high-dose region, 3% global difference in the low-dose region, 3 mm distance to agreement in the penumbra, and a gamma pass rate of >95% for 3%/3 mm criteria. Acuros was able to accurately calculate dose through cork and bone-equivalent heterogeneities. AAA was able to accurately calculate dose through the bone-equivalent heterogeneity but did not pass within the recommended criteria for the cork heterogeneity. For the 2.5 MV imaging beam, both the AAA and Acuros algorithms provide calculated doses that agree with measured results well within the 20% criteria for imaging beams recommended by AAPM TG-180.

Download full-text PDF

Source
http://dx.doi.org/10.1002/acm2.12756DOI Listing
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC6909176PMC
December 2019
1 Read

Publication Analysis

Top Keywords

treatment planning
12
imaging beam
12
planning system
8
calculated doses
8
accurately calculate
8
25 mv imaging
8
dose 25 mv
8
commercial treatment
8
calculate dose
8
beam
6
recommended criteria
4
subsequently verified
4
criteria cork
4
algorithms subsequently
4
cork heterogeneity
4
mppg guidelines
4
guidelines calculated
4
agreement penumbra
4
acuros accurately
4
modeling algorithms
4

Similar Publications