Trends in radiation dose and image quality for pediatric patients with a multidetector CT and a third-generation dual-source dual-energy CT.

Authors:
Cecilia Lanza
Cecilia Lanza
Azienda Ospedaliero-Universitaria Ospedali Riuniti
Italy
Alessandra Borgheresi
Alessandra Borgheresi
Santa Maria Nuova Hospital
Italy
Stefania Maggi
Stefania Maggi
Institute of Neuroscience
Italy
Andrea Giovagnoni
Andrea Giovagnoni
Università Politecnica delle Marche
Italy

Radiol Med 2019 Apr 19. Epub 2019 Apr 19.

University Politecnica delle Marche, Via Tronto 10/A, 60126, Ancona, Italy.

Aim: To provide an overview on dose reduction and image quality after the installation of a third-generation dual-source CT (dsCT) in a Pediatric Radiology Department.

Materials And Methods: We included pediatric patients (< 20 years old) undergoing CT for oncological staging (neck, chest and abdomen) or low-dose chest CT for lung diseases. Each of these two groups were further divided in two age groups (≤ or > 10 years old) including patients scanned in the same period of two consecutive years, in 2017 with a 16-row LightSpeed CT (GE Healthcare) or in 2018 with a Somatom Force dsCT (Siemens Healthineers). Technical parameters such as kVp, mAs, slice thickness, exposure times and dose indicators were retrieved and compared. Image quality was evaluated in consensus by two radiologists on a five-point semiquantitative scale. Nonparametric tests were used.

Results: In oncological patients, significantly lower kVp and tube current with better image quality were achieved with the dsCT. Radiation dose (total DLP) was 5-6 times lower with dsCT, thanks also to virtual non-contrast images. In low-dose chest CT, the frequent use of tin filter required higher tube current; a total DLP 3 times lower was achieved with dsCT in patients ≤ 10 years old. The image quality was better with the dsCT in low-dose chest CT protocols.

Conclusion: The third-generation dsCT provides high-quality images with reduced motion artifacts at lower dose.

Download full-text PDF

Source
http://link.springer.com/10.1007/s11547-019-01037-5
Publisher Site
http://dx.doi.org/10.1007/s11547-019-01037-5DOI Listing
April 2019
1 Read

Publication Analysis

Top Keywords

image quality
20
pediatric patients
8
tube current
8
third-generation dual-source
8
low-dose chest
8
times lower
8
radiation dose
8
total dlp
8
achieved dsct
8
dsct
7
quality
5
dose
5
image
5
lower
4
usedresults oncological
4
nonparametric tests
4
patients lower
4
tests usedresults
4
oncological patients
4
current better
4

References

(Supplied by CrossRef)
Article in World J Radiol
H Ferris et al.
World J Radiol 2016
Article in Radiology
DP Frush et al.
Radiology 1998
Article in The Lancet
MS Pearce et al.
The Lancet 2012
Article in J Natl Cancer Inst
JM Meulepas et al.
J Natl Cancer Inst 2018
Article in Radiol Med
P Toma et al.
Radiol Med 2017
Article in Eur J Radiol
B Bodelle et al.
Eur J Radiol 2017
Article in Korean J Radiol
HW Goo et al.
Korean J Radiol 2018
Article in Eur Radiol
TG Flohr et al.
Eur Radiol 2006
Article in Invest Radiol
MM Lell et al.
Invest Radiol 2011
Article in Eur Radiol
TR Johnson et al.
Eur Radiol 2007
Article in Radiology
MJ Siegel et al.
Radiology 2019

Similar Publications