First comparison of performances between the new whole-body cadmium-zinc-telluride SPECT-CT camera and a dedicated cardiac CZT camera for myocardial perfusion imaging: Analysis of phantom and patients.

J Nucl Cardiol 2019 Apr 8. Epub 2019 Apr 8.

Nuclear Medicine Department, IRIS, Hôpital Privé Le Bois, Lille, France.

Background: Dedicated cardiac Cadmium-zinc-telluride (CZT) cameras show superior performances compared with Anger systems, particularly in terms of spatial resolution and count sensitivity. This study evaluated the performances of a new polyvalent whole body CZT camera (DNM 670CZT) compared with a cardiac dedicated CZT camera (DNM 530c) for myocardial perfusion SPECT.

Methods: The spatial resolution was evaluated with three linear sources filled with Tc. We used a cardiac phantom to evaluate count sensitivity, sharpness index, contrast-to-noise ratio, wall thickness, non-uniformity index, perfusion scores and ventricle volumes for both cameras. The impact of matrix size, and acquisition time was investigated. Concordance between the two cameras was evaluated in patients using QPS/QGS software for quantitative segmental perfusion, motion and thickness scores.

Results: The spatial resolution was identical with the two cameras. Count sensitivity of the DNM 670CZT was twofold lower compared with the DNM 530c, leading to lower sharpness index and contrast-to-noise ratio. The wall thickness and the myocardial volumes were similar. Visual and quantitative assessments of the perfusion patterns have shown a good concordance of the two cameras on phantoms and in patients.

Conclusion: This study demonstrated the feasibility of myocardial perfusion SPECT imaging using the new whole-body DNM 670CZT camera.

Download full-text PDF

Source
http://link.springer.com/10.1007/s12350-019-01702-2
Publisher Site
http://dx.doi.org/10.1007/s12350-019-01702-2DOI Listing

Still can't find the full text of the article?

We can help you send a request to the authors directly.
April 2019
6 Reads

Publication Analysis

Top Keywords

count sensitivity
12
spatial resolution
12
myocardial perfusion
12
czt camera
12
dnm 670czt
12
wall thickness
8
concordance cameras
8
dedicated cardiac
8
sharpness contrast-to-noise
8
ratio wall
8
dnm 530c
8
camera dnm
8
contrast-to-noise ratio
8
perfusion
6
cameras
5
dnm
5
camera
5
evaluate count
4
compared dnm
4
phantom evaluate
4

References

(Supplied by CrossRef)
Article in EJNMMI Phys.
T Blaire et al.
EJNMMI Phys. 2016
Article in J Nucl Cardiol.
A Bailliez et al.
J Nucl Cardiol. 2014
Article in J Nucl Med.
SS Gambhir et al.
J Nucl Med. 2009
Article in Phys Med Biol.
K Erlandsson et al.
Phys Med Biol. 2009
Article in Eur J Nucl Med Mol Imaging.
D Bellevre et al.
Eur J Nucl Med Mol Imaging. 2015
Article in Eur J Nucl Med Mol Imaging.
A Verger et al.
Eur J Nucl Med Mol Imaging. 2013
Article in Ann Nucl Med.
AS Kroiss et al.
Ann Nucl Med. 2017
Article in Eur J Nucl Med Mol Imaging.
M Mouden et al.
Eur J Nucl Med Mol Imaging. 2012
Article in J Nucl Cardiol.
WL Duvall et al.
J Nucl Cardiol. 2011
Article in Ann Nucl Med.
Y Takahashi et al.
Ann Nucl Med. 2013
Article in J Nucl Med.
L Imbert et al.
J Nucl Med. 2012

Similar Publications