Bone resorption triggered by high radial stress: The mechanism of screw loosening in plate fixation of long bone fractures.

Authors:
William W Lu
William W Lu
The University of Hong Kong
Hong Kong
Bin Chen
Bin Chen
Chongqing Key Laboratory of Vector Insects; Institute of Entomology and Molecular Biology
China

J Orthop Res 2019 Mar 25. Epub 2019 Mar 25.

Department of Orthopaedics and Traumatology, Queen Mary Hospital, the University of Hong Kong, Pok Fu Lam, Hong Kong.

Screw loosening is a common complication in plate fixation. However, the underlying mechanism is unclear. This study investigated screw loosening mechanisms by finite element analysis (FEA) simulation and clinical X-ray feature analysis. Two FEA models incorporated bone heterogeneity and orthotropy, representing fracture fixation using dynamic compression plate (DCP) and locking compression plate (LCP), were developed. These models were used to examine the volume of bone exceeding a certain stress value around each screw under physiologically-relevant loading conditions. These damaged bone was then separated and compared by the axial stress and radial stress of each screw. In addition, features of patients' X-ray images showing screw loosening were analyzed to validate the loosening features simulated by the models. The FEA study showed that more damaged bone was found at the central two screws which gradually decreased toward the two end screws in all groups. More bone was damaged by the radial stress of each screw than by the axial stress. The radiological analysis of screw loosening showed that bone loss occurred at the screw closest to the fracture line first then subsequent bone loss at the screws further away from the fracture line occurred. This study found that the two screws nearest to the fracture line are more vulnerable to loosening. The radial stress of the screw plays a larger role in screw loosening than the axial stress. Bone resorption triggered by the high radial stress of screws is indicated as the mechanism of screw loosening in the diaphyseal plate fixation. © 2019 Orthopaedic Research Society. Published by Wiley Periodicals, Inc. J Orthop Res.

Download full-text PDF

Source
http://dx.doi.org/10.1002/jor.24286DOI Listing
March 2019
2 Reads

Article Mentions


Provided by Crossref Event Data
twitter
Twitter:
April 9, 2019, 11:34 am EST

Publication Analysis

Top Keywords

screw loosening
28
radial stress
20
stress screw
16
plate fixation
12
screw
12
axial stress
12
bone
10
stress
9
loosening
9
bone resorption
8
bone loss
8
compression plate
8
mechanism screw
8
damaged bone
8
resorption triggered
8
triggered high
8
high radial
8
analysis fea
8
screws
5
plate
5

References

(Supplied by CrossRef)
Evaluation of threshold stress for bone resorption around screws based on in vivo strain measurement of miniplate
Sugiura T et al.
J Musculoskelet Neuronal Interact 2000
Stress shielding and bone resorption in THA: clinical versus computer‐simulation studies
Huiskes R et al.
Acta Orthop Belg 1993
Dynamic simulations of cancellous bone resorption around orthopaedic fixative implants
Gefen A et al.
Eng Med Biol Soc Proc Ann Int Conf 2001

Similar Publications