4D liver tumor localization using cone-beam projections and a biomechanical model.

Radiother Oncol 2019 04 14;133:183-192. Epub 2018 Nov 14.

Department of Radiation Oncology, UT Southwestern Medical Center, Dallas, USA.

Purpose: To improve the accuracy of liver tumor localization, this study tests a biomechanical modeling-guided liver cone-beam CT (CBCT) estimation (Bio-CBCT-est) technique, which generates new CBCTs by deforming a prior high-quality CT or CBCT image using deformation vector fields (DVFs). The DVFs can be used to propagate tumor contours from the prior image to new CBCTs for automatic 4D tumor localization.

Methods/materials: To solve the DVFs, the Bio-CBCT-est technique employs an iterative scheme that alternates between intensity-driven 2D-3D deformation and biomechanical modeling-guided DVF regularization and optimization. The 2D-3D deformation step solves DVFs by matching digitally reconstructed radiographs of the 3D deformed prior image to 2D phase-sorted on-board projections according to imaging intensities. This step's accuracy is limited at low-contrast intra-liver regions without sufficient intensity variations. To boost the DVF accuracy in these regions, we use the intensity-driven DVFs solved at higher-contrast liver boundaries to fine-tune the intra-liver DVFs by finite element analysis-based biomechanical modeling. We evaluated Bio-CBCT-est's accuracy with seven liver cancer patient cases. For each patient, we simulated 4D cone-beam projections from 4D-CT images, and used these projections for Bio-CBCT-est based image estimations. After Bio-CBCT-est, the DVF-propagated liver tumor/cyst contours were quantitatively compared with the manual contours on the original 4D-CT 'reference' images, using the DICE similarity index, the center-of-mass-error (COME), the Hausdorff distance (HD) and the voxel-wise cross-correlation (CC) metrics. In addition to simulation, we also performed a preliminary study to qualitatively evaluate the Bio-CBCT-est technique via clinically acquired cone beam projections. A quantitative study using an in-house deformable liver phantom was also performed.

Results: Using 20 projections for image estimation, the average (±s.d.) DICE index increased from 0.48 ± 0.13 (by 2D-3D deformation) to 0.77 ± 0.08 (by Bio-CBCT-est), the average COME decreased from 7.7 ± 1.5 mm to 2.2 ± 1.2 mm, the average HD decreased from 10.6 ± 2.2 mm to 5.9 ± 2.0 mm, and the average CC increased from -0.004 ± 0.216 to 0.422 ± 0.206. The tumor/cyst trajectory solved by Bio-CBCT-est matched well with that manually obtained from 4D-CT reference images.

Conclusions: Bio-CBCT-est substantially improves the accuracy of 4D liver tumor localization via cone-beam projections and a biomechanical model.

Download full-text PDF

Source
https://linkinghub.elsevier.com/retrieve/pii/S01678140183358
Publisher Site
http://dx.doi.org/10.1016/j.radonc.2018.10.040DOI Listing
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC6445758PMC
April 2019
55 Reads
4.363 Impact Factor

Publication Analysis

Top Keywords

cone-beam projections
12
accuracy liver
12
2d-3d deformation
12
bio-cbct-est technique
12
tumor localization
12
liver tumor
12
bio-cbct-est
8
biomechanical modeling-guided
8
liver
8
biomechanical model
8
average decreased
8
prior image
8
localization cone-beam
8
projections biomechanical
8
projections
7
dvfs
6
biomechanical
5
accuracy
5
image
5
bio-cbct-est dvf-propagated
4

Similar Publications