Accuracy assessment of a novel image-free handheld robot for Total Knee Arthroplasty in a cadaveric study.

Authors:
Rahul Khare
Rahul Khare
Northwestern University
Branislav Jaramaz
Branislav Jaramaz
The Western Pennsylvania Hospital
United States
Brian Hamlin
Brian Hamlin
The Bone & Joint Center
David Mayman
David Mayman
Hospital for Special Surgery
United States
Jeff Headrick
Jeff Headrick
Center for Orthopedic Surgery
Charter Township of Clinton | United States

Comput Assist Surg (Abingdon) 2018 12;23(1):14-20

l Orthopedic Surgeon, Scott Orthopedic Center , Huntington , WV , USA.

Surgical navigation has been shown to improve the accuracy of bone preparation and limb alignment in total knee arthroplasty (TKA). Previous work has shown the effectiveness of various types of navigation systems. Here, for the first time, we assessed the accuracy of a novel imageless semiautonomous handheld robotic sculpting system in performing bone resection and preparation in TKA using cadaveric specimens. In this study, we compared the planned and final implant placement in 18 cadaveric specimens undergoing TKA using the new tool. Eight surgeons carried out the procedures using three types of implant designs. A quantitative analysis was performed to determine the translational, angular, and rotational differences between the planned and achieved positions of the implants. The mean femoral flexion, varus/valgus, and rotational error was -2.0°, -0.1°, and -0.5°, respectively. The mean tibial posterior slope, and varus/valgus error was -0.2°, and -0.2°, respectively. We obtained higher flexion errors for the femoral implant when using cut-guides as compared to using a bur for cutting the bones. The image-free robotic sculpting tool achieved accurate implementation of the surgical plan with small errors in implant placement. Future studies will focus on determining how well the accurate implant placement translates into a clinical and functional benefit for the patient.

Download full-text PDF

Source
https://www.tandfonline.com/doi/full/10.1080/24699322.2018.1
Publisher Site
http://dx.doi.org/10.1080/24699322.2018.1519038DOI Listing
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC6309369PMC

Still can't find the full text of the article?

We can help you send a request to the authors directly.
December 2018
6 Reads

Publication Analysis

Top Keywords

implant placement
12
total knee
8
cadaveric specimens
8
knee arthroplasty
8
robotic sculpting
8
implant
5
positions implants
4
achieved positions
4
implants femoral
4
differences planned
4
angular rotational
4
rotational differences
4
planned achieved
4
flexion varus/valgus
4
-01° -05°
4
-05° tibial
4
tibial posterior
4
-20° -01°
4
error -20°
4
translational angular
4

References

(Supplied by CrossRef)
Article in Acta Orthop Belg
Nizard R et al.
Acta Orthop Belg 2002
Article in Clin Orthop
Ritter MA et al.
Clin Orthop 1994
Article in Bone Joint J
Jeffery RS et al.
Bone Joint J 1991