Commissioning of a dedicated commercial Co-60 total body irradiation unit.

J Appl Clin Med Phys 2018 May 11;19(3):131-141. Epub 2018 Mar 11.

Karmanos Cancer Center, Gershenson R.O.C., Detroit, MI, USA.

We describe the commissioning of the first dedicated commercial total body irradiation (TBI) unit in clinical operation. The Best Theratronics GammaBeam 500 is a Co-60 teletherapy unit with extended field size and imaging capabilities. Radiation safety, mechanical and imaging systems, and radiation output are characterized. Beam data collection, calibration, and external dosimetric validation are described. All radiation safety and mechanical tests satisfied relevant requirements and measured dose distributions meet recommendations of American Association of Physicists in Medicine (AAPM) Report #17. At a typical treatment distance, the dose rate in free space per unit source activity using the thick flattening filter is 1.53 × 10 cGy*min *Ci . With a 14,000 Ci source, the resulting dose rate at the midplane of a typical patient is approximately 17 and 30 cGy/min using the thick and thin flattening filters, respectively, using the maximum source to couch distance. The maximum useful field size, defined by the 90% isodose line, at this location is 225 × 78 cm with field flatness within 5% over the central 178 × 73 cm. Measured output agreed with external validation within 0.5%. End-to-end testing was performed in a modified Rando phantom. In-house MATLAB software was developed to calculate patient-specific dose distributions using DOSXYZnrc, and fabricate custom 3D-printed forms for creating patient-specific lung blocks. End-to-end OSLD and diode measurements both with and without lung blocks agreed with Monte Carlo calculated doses to within 5% and in-phantom film measurements validated dose distribution uniformity. Custom lung block transmission measurements agree well with design criteria and provide clinically favorable dose distributions within the lungs. Block placement is easily facilitated using the flat panel imaging system with an exposure time of 0.01 min. In conclusion, a novel dedicated TBI unit has been commissioned and clinically implemented. Its mechanical, dosimetric, and imaging capabilities are suitable to provide state-of-the-art TBI for patients as large as 220 cm in height and 78 cm in width.

Download full-text PDF

Source
http://doi.wiley.com/10.1002/acm2.12309
Publisher Site
http://dx.doi.org/10.1002/acm2.12309DOI Listing
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC5978703PMC
May 2018
6 Reads

Publication Analysis

Top Keywords

dose distributions
12
body irradiation
8
total body
8
radiation safety
8
imaging capabilities
8
tbi unit
8
lung blocks
8
field size
8
dose rate
8
safety mechanical
8
commissioning dedicated
8
dedicated commercial
8
dose
6
unit
5
source dose
4
imaging system
4
14000 source
4
custom 3d-printed
4
agree well
4
153 × 10 cgy*min
4

References

(Supplied by CrossRef)
The physical aspects of total and half body photon irradiation
Dyk et al.
Task Group 1986
Total body irradiation with a sweeping beam
Pla et al.
Int J Radiat Oncol Biol Phys 1983
Cobalt-60 therapy unit for large field irradiation
Leung et al.
Int J Radiat Oncol Biol Phys 1981
Central axis depth dose data for use in radiotherapy
Cohen et al.
Brit J Radiol 1972

Similar Publications