Influence of thread shape and inclination on the biomechanical behaviour of plateau implant systems.

Authors:
elisabetta zanetti
elisabetta zanetti
Università degli Studi di Perugia
Assistant Professor
biomechanics, biomaterials
Perugia, Umbria | Italy

Dent Mater 2018 03 1;34(3):460-469. Epub 2018 Feb 1.

DIMEAS, Politecnico di Torino, Cso Duca degli Abruzzi, 24, 10129 Torino, Italy. Electronic address:

Objective: To assess the influence of implant thread shape and inclination on the mechanical behaviour of bone-implant systems. The study assesses which factors influence the initial and full osseointegration stages.

Methods: Point clouds of the original implant were created using a non-contact reverse engineering technique. A 3D tessellated surface was created using Geomagic Studio software. From cross-section curves, generated by intersecting the tessellated model and cutting-planes, a 3D parametric CAD model was created using SolidWorks 2017. By the permutation of three thread shapes (rectangular, 30° trapezoidal, 45° trapezoidal) and three thread inclinations (0°, 3° or 6°), nine geometric configurations were obtained. Two different osseointegration stages were analysed: the initial osseointegration and a full osseointegration. In total, 18 different FE models were analysed and two load conditions were applied to each model. The mechanical behaviour of the models was analysed by Finite Element (FE) Analysis using ANSYS v. 17.0. Static linear analyses were also carried out.

Results: ANOVA was used to assess the influence of each factor. Models with a rectangular thread and 6° inclination provided the best results and reduced displacement in the initial osseointegration stages up to 4.58%. This configuration also reduced equivalent VM stress peaks up to 54%. The same effect was confirmed for the full osseointegration stage, where 6° inclination reduced stress peaks by up to 62%.

Significance: The FE analysis confirmed the beneficial effect of thread inclination, reducing the displacement in immediate post-operative conditions and equivalent VM stress peaks. Thread shape does not significantly influence the mechanical behaviour of bone-implant systems but contributes to reducing stress peaks in the trabecular bone in both the initial and full osseointegration stages.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.dental.2018.01.012DOI Listing
March 2018
30 Reads
3.769 Impact Factor

Publication Analysis

Top Keywords

full osseointegration
16
stress peaks
16
mechanical behaviour
12
osseointegration stages
12
thread shape
12
behaviour bone-implant
8
assess influence
8
initial full
8
initial osseointegration
8
bone-implant systems
8
shape inclination
8
equivalent stress
8
models analysed
8
6° inclination
8
three thread
8
osseointegration
7
thread
6
inclination
5
influence
5
model mechanical
4

Similar Publications