Fully automated chemiluminescence detection using an electrified-Lab-on-a-Disc (eLoaD) platform.

Authors:
Niamh A Kilcawley
Niamh A Kilcawley
Biomedical Diagnostics Institute
Dr. Jens Ducree, Dr. rer. nat. habil. Dipl. Phys.
Dr. Jens Ducree, Dr. rer. nat. habil. Dipl. Phys.
Fraunhofer Project Centre at Dublin City University
Professor (Full)
microfluidics, Lab-on-a-Chip, hydrodynanmics, business development, project management, organisational leadership
Glasnevin, Dublin 9 | Ireland
Dario Mager
Dario Mager
University of Freiburg

Lab Chip 2016 10;16(20):4002-4011

Institute of Microstructure Technology, Karlsruhe Institute of Technology, Hermann-von-Helmholtz-Platz 1, Eggenstein-Leopoldshafen 76344, Germany.

Typical Lab-on-a-Disc (LoaD) platforms cannot make a continuous measurement while the disc is spinning; this drawback means that the disc usually must be stopped and aligned with a sensor. This can result in measurement errors in time-dependent assays along with inaccuracies due to liquid displacement and bubble formation in the absence of a stabilising centrifugal field. This paper presents a novel concept for a wirelessly electrified-Lab-on-a-Disc (eLoaD) platform that allows continuous measurement of experimental parameters while the disc is spinning. This platform incorporates all the components needed for measurement within the rotating frame of reference, and bidirectional transmission of data outside this reference frame, thus allowing for online measurement independent of the rotation of the disc. The eLoaD platform is conceived in a modular manner whereby an interchangeable and non-disposable 'Application Disc' can be fitted to the eLoaD platform and so the system can be adapted for a range of optical, electrochemical and other measurement types. As an application example, optical readout, using the Application Disc fitted with a silicon photomultiplier, is demonstrated using a tagged chemiluminescent antibody, which is commonly used, for instance, in ELISA assays. The precision of the eLoaD platform is >94%, while its accuracy, when compared to a commercial benchtop luminometer, is higher than 96%. The modular design of this platform will permit extension of this technology to many other LoaD applications.

Download full-text PDF

Source
http://dx.doi.org/10.1039/c6lc00973eDOI Listing
October 2016
27 Reads

Publication Analysis

Top Keywords

eload platform
20
electrified-lab-on-a-disc eload
8
disc spinning
8
continuous measurement
8
platform
7
measurement
6
disc
5
eload
5
modular manner
4
manner interchangeable
4
platform conceived
4
disc eload
4
interchangeable non-disposable
4
conceived modular
4
disc' fitted
4
adapted range
4
range optical
4
optical electrochemical
4
electrochemical measurement
4
system adapted
4

References

(Supplied by CrossRef)
Article in J. Lab. Autom.
Kong et al.
J. Lab. Autom. 2016
Article in J. Micromech. Microeng.
Ducrée et al.
J. Micromech. Microeng. 2007
Article in Chem. Soc. Rev.
Strohmeier et al.
Chem. Soc. Rev. 2015
Article in PLoS One
Kinahan et al.
PLoS One 2016
Article in Sens. Actuators, A
Kinahan et al.
Sens. Actuators, A 2014
Article in Lab Chip
Gorkin et al.
Lab Chip 2010
Article in Anal. Chem.
Kim et al.
Anal. Chem. 2014
Article in Biosens. Bioelectron.
Nwankire et al.
Biosens. Bioelectron. 2015
Article in Anal. Chem.
Park et al.
Anal. Chem. 2014
Article in Anal. Chem.
Hwang et al.
Anal. Chem. 2013
Article in Sens. Actuators, B
Amasia et al.
Sens. Actuators, B 2012

Similar Publications