    Fast Megavoltage Computed Tomography: A Rapid Imaging Method for Total Body or Marrow Irradiation in Helical Tomotherapy.
    Int J Radiat Oncol Biol Phys 2016 Nov 6;96(3):688-95. Epub 2016 Jul 6.
    Masonic Cancer Center, University of Minnesota, Minneapolis, Minnesota; Department of Therapeutic Radiology, University of Minnesota, Minneapolis, Minnesota; Department of Radiation Oncology and Beckman Research Institute, City of Hope, Duarte, California. Electronic address:
    Purpose: Megavoltage computed tomographic (MVCT) imaging has been widely used for the 3-dimensional (3-D) setup of patients treated with helical tomotherapy (HT). One drawback of MVCT is its very long imaging time, the result of slow couch speeds of approximately 1 mm/s, which can be difficult for the patient to tolerate. We sought to develop an MVCT imaging method allowing faster couch speeds and to assess its accuracy for image guidance for HT.

    Methods And Materials: Three cadavers were scanned 4 times with couch speeds of 1, 2, 3, and 4 mm/s. The resulting MVCT images were reconstructed using an iterative reconstruction (IR) algorithm with a penalty term of total variation and with a conventional filtered back projection (FBP) algorithm. The MVCT images were registered with kilovoltage CT images, and the registration errors from the 2 reconstruction algorithms were compared. This fast MVCT imaging was tested in 3 cases of total marrow irradiation as a clinical trial.

    Results: The 3-D registration errors of the MVCT images reconstructed with the IR algorithm were smaller than the errors of images reconstructed with the FBP algorithm at fast couch speeds (2, 3, 4 mm/s). The scan time and imaging dose at a speed of 4 mm/s were reduced to 30% of those from a conventional coarse mode scan. For the patient imaging, faster MVCT (3 mm/s couch speed) scanning reduced the imaging time and still generated images useful for anatomic registration.

    Conclusions: Fast MVCT with the IR algorithm is clinically feasible for large 3-D target localization, which may reduce the overall time for the treatment procedure. This technique may also be useful for calculating daily dose distributions or organ motion analyses in HT treatment over a wide area. Automated integration of this imaging is at least needed to further assess its clinical benefits.

    Similar Publications

    Multi-institutional feasibility study of a fast patient localization method in total marrow irradiation with helical tomotherapy: a global health initiative by the international consortium of total marrow irradiation.
    Int J Radiat Oncol Biol Phys 2015 Jan 30;91(1):30-8. Epub 2014 Oct 30.
    Masonic Cancer Center, University of Minnesota, Minneapolis, Minnesota; Department of Therapeutic Radiology, University of Minnesota, Minneapolis, Minnesota. Electronic address:
    Purpose: To develop, characterize, and implement a fast patient localization method for total marrow irradiation.

    Methods And Materials: Topographic images were acquired using megavoltage computed tomography (MVCT) detector data by delivering static orthogonal beams while the couch traversed through the gantry. Geometric and detector response corrections were performed to generate a megavoltage topogram (MVtopo). Read More
    SU-E-J-38: Fast Patient Localization Method in Total Marrow Irradiation with Helical Tomotherapy.
    Med Phys 2013 Jun;40(6Part7):158
    University of Minnesota, Minneapolis, MN.
    Purpose: Reducing imaging and treatment times is a crucial issue in total marrow irradiation (TMI) with helical tomotherapy. The purpose of this study was to develop the fast patient localization method for TMI patients with helical tomotherapy.

    Methods: The topographic images were acquired by getting the megavoltage CT (MVCT) after running calibration procedures with the couch traversing through the gantry at static orthogonal beam angles. Read More
    Quantitative characterization of tomotherapy MVCT dosimetry.
    Med Dosim 2013 1;38(3):280-6. Epub 2013 Apr 1.
    21st Century Oncology, Madison, WI 53719, USA.
    Megavoltage computed tomography (MVCT) is used as image guidance for patient setup in almost every tomotherapy treatment. Frequent use of ionizing radiation for image guidance has raised concern of imaging dose. The purpose of this work is to quantify and characterize tomotherapy MVCT dosimetry. Read More
    Estimation of setup uncertainty using planar and MVCT imaging for gynecologic malignancies.
    Int J Radiat Oncol Biol Phys 2008 Aug 4;71(5):1511-7. Epub 2008 Jun 4.
    Department of Radiation Oncology, Washington University School of Medicine, 4921 Parkview Place, St. Louis, MO 63110-1093, USA.
    Purpose: This prospective study investigates gynecologic malignancy online treatment setup error corrections using planar kilovoltage/megavoltage (KV/MV) imaging and helical MV computed tomography (MVCT) imaging.

    Methods And Materials: Twenty patients were divided into two groups. The first group (10 patients) was imaged and treated using a conventional linear accelerator (LINAC) with image-guidance capabilities, whereas the second group (10 patients) was treated using tomotherapy with MVCT capabilities. Read More