Agreement between clinical history method, Orbscan IIz, and Pentacam in estimating corneal power after myopic excimer laser surgery.

Authors:
Kaevalin Lekhanont Manachai Nonpassopon Khemruetai Wannarosapark Varintorn Chuckpaiwong

PLoS One 2015 8;10(4):e0123729. Epub 2015 Apr 8.

Department of Ophthalmology, Ramathibodi Hospital, Mahidol University, Bangkok, Thailand.

The purpose of this study was to investigate the agreement between the clinical history method (CHM), Orbscan IIz, and Pentacam in estimating corneal power after myopic excimer laser surgery. Fifty five patients who had myopic LASIK/PRK were recruited into this study. One eye of each patient was randomly selected by a computer-generated process. At 6 months after surgery, postoperative corneal power was calculated from the CHM, Orbscan IIz total optical power at the 3.0 and 4.0 mm zones, and Pentacam equivalent keratometric readings (EKRs) at 3.0, 4.0, and 4.5 mm. Statistical analyses included multilevel models, Pearson's correlation test, and Bland-Altman plots. The Orbscan IIz 3.0-mm and 4.0 mm total optical power, and Pentacam 3.0-mm, 4.0-mm, and 4.5-mm EKR values had strong linear positive correlations with the CHM values (r = 0.90-0.94, P = <0.001, for all comparisons, Pearson's correlation). However, only Pentacam 3.0-mm EKR was not statistically different from CHM (P = 0.17, multilevel models). The mean 3.0- and 4.0-mm total optical powers of the Orbscan IIz were significantly flatter than the values derived from CHM, while the average EKRs of the Pentacam at 4.0 and 4.5 mm were significantly steeper. The mean Orbscan IIz 3.0-mm total optical power was the lowest keratometric reading compared to the other 5 values. Large 95% LoA was observed between each of these values, particularly EKRs, and those obtained with the CHM. The width of the 95% LoA was narrowest for Orbscan IIz 3.0-mm total optical power. In conclusion, the keratometric values extracted from these 3 methods were disparate, either because of a statistically significant difference in the mean values or moderate agreement between them. Therefore, they are not considered equivalent and cannot be used interchangeably.

Download full-text PDF

Source
http://journals.plos.org/plosone/article?id=10.1371/journal.pone.0123729PLOS
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC4390196PMC
March 2016

Publication Analysis

Top Keywords

orbscan iiz
16
corneal power
12
total optical
8
optical power
8
iiz pentacam
8
power myopic
8
pentacam estimating
8
myopic excimer
8
chm orbscan
8
agreement clinical
8
clinical history
8
laser surgery
8
excimer laser
8
history method
8
estimating corneal
8
power
5
readings ekrs
4
keratometric readings
4
equivalent keratometric
4
ekrs statistical
4

Keyword Occurance

Similar Publications

Changes of angle Kappa and corneal morphology changes in myopic patients after SubBowmanKeratomileusis.

Authors:
Yi Ding Zhipei Zhang Xiaoshan Min Xiaobo Xia Shengfa Hu

Zhong Nan Da Xue Xue Bao Yi Xue Ban 2021 Feb;46(2):162-168

Department of Ophthalmology, Xiangya Hospital, Central South University, Changsha 410008.

Objectives: To investigate angle Kappa and diopter distribution in myopic patients and the changes of angle Kappa and corneal morphology after Sub-Bowman-Keratomileusis (SBK), and to analyze the effects of the surgery on corneal morphologic changes and the patients' near fixation characteristics.

Methods: The clinical data of 134 myopic patients (268 eyes) undergoing SBK from August 2015 to August 2016 were retrospectively analyzed. Angle Kappa, corneal curvature in the central corneal region of 3 mm, and post-corneal Diff value were measured by Orbscan IIz Corneal Topography System before operation, 1 month and 6 months after operation. Read More

View Article and Full-Text PDF
February 2021
Similar Publications

Comparison of central corneal thickness measurement by scanning slit topography, infrared, and ultrasound pachymetry in normal and post-LASIK eyes.

Authors:
Wing-Cheung Ho Philip Tsze-Ho Lam Thomas Yee-Hang Chiu Mandy Ching-Man Yim Fion Tung-Ching Lau

Int Ophthalmol 2020 Nov 3;40(11):2913-2921. Epub 2020 Jul 3.

Eye Centre, Hong Kong Baptist Hospital, 222 Waterloo Road, Kowloon Tong, Hong Kong, People's Republic of China.

Objectives: To compare central corneal thickness (CCT) measurements by scanning slit topography (SST), infrared pachymetry (IRP), and ultrasound pachymetry (USP), and their agreement in normal and post-laser in situ keratomileusis (LASIK) eyes.

Methods: Sixty normal and 35 post-LASIK subjects were recruited. Only one eye from each subject was analyzed. Read More

View Article and Full-Text PDF
November 2020
Similar Publications

Adjustment formulae to improve the correlation of white-to-white measurement with direct measurement of the ciliary sulcus diameter by ultrasound biomicroscopy.

Authors:
Seyed Javad Hashemian Majid Mohebbi Mehdi Yaseri Mohammad Ebrahim Jafari Shadrokh Nabili Seyed Mahyar Hashemian Mahsa Sadat Hashemian

J Curr Ophthalmol 2018 Sep 16;30(3):217-222. Epub 2017 Dec 16.

School of Pharmacy, International Campus, Tehran University of Medical Sciences, Tehran, Iran.

Purpose: This study evaluates the correlation between horizontal white-to-white (WTW) distance using Caliper and Orbscan IIz with the ciliary sulcus diameter measured by high frequency ultrasound biomicroscopy (UBM) and presents an adjustment formula to improve the correlation.

Methods: We measured horizontal sulcus-to-sulcus (STS) dimension of 273 right eyes of 273 high myopic patients with 35 MHz UBM and horizontal WTW using Orbscan IIz and Caliper. Mean WTW diameter, differences, and the correlation of measurement methods were evaluated. Read More

View Article and Full-Text PDF
September 2018
Similar Publications

Improving precision for detecting change in the shape of the cornea in patients with keratoconus.

Authors:
Matthias Brunner Gabriela Czanner Riccardo Vinciguerra Vito Romano Sajjad Ahmad Mark Batterbury Claire Britten Colin E Willoughby Stephen B Kaye

Sci Rep 2018 08 17;8(1):12345. Epub 2018 Aug 17.

Department of Corneal and External Eye Diseases, St. Paul's Eye Unit, Royal Liverpool University Hospital, Liverpool, United Kingdom.

To investigate a method for precision analysis to discriminate true corneal change from measurement imprecision in keratoconus (KC). Thirty patients with KC and 30 healthy controls were included. Coefficients of repeatability and limits of agreement (LOA) were compared using multiple measurements for inter-observer and inter-device agreement with the Pentacam HR, Orbscan IIz, and Tomey Casia SS-1000. Read More

View Article and Full-Text PDF
August 2018
Similar Publications

Analysis of cases and accuracy of 3 risk scoring systems in predicting ectasia after laser in situ keratomileusis.

Authors:
Cordelia Chan Alain Saad J Bradley Randleman Mona Harissi-Dagher Daniel Chua Mujtaba Qazi Jean-Jacques Saragoussi Rohit Shetty Jean-Marc Ancel Robert Ang Dan Z Reinstein Damien Gatinel

J Cataract Refract Surg 2018 Aug;44(8):979-992

From Singapore National Eye Center (Chan, Chua) and Eye Surgeons @ Novena (Chan), Singapore, Singapore; the Rothschild Foundation (Saad, Gatinel), Clinique Lamartine (Ancel), and the Centre D'Ophtalmologie (Saragoussi), Paris, France; the American University of Beirut (Saad), Beirut, Lebanon; Roski Eye Institute (Randleman), University of Southern California, Los Angeles, California, and Pepose Vision Institute (Qazi), Chesterfield, Missouri, USA; the Departement d'Ophtalmologie (Harissi-Dagher), Université de Montréal, Montreal, Quebec, Canada; Narayana Nethralaya Hospital (Shetty), Bangalore, India; Asian Eye Institute (Ang), Manila, Philippines; London Vision Clinic (Reinstein), London, United Kingdom. Electronic address:

Purpose: To identify risk factors for ectasia after laser in situ keratomileusis (LASIK) by comparing the accuracy of the Ectasia Risk Score System (ERSS), Screening Corneal Objective Risk of Ectasia (SCORE) Analyzer, and percentage of tissue altered (PTA) in predicting the occurrence of ectasia.

Setting: Multiple centers in 8 countries.

Design: Retrospective case series. Read More

View Article and Full-Text PDF
August 2018
Similar Publications
© 2021 PubFacts.
  • About PubFacts
  • Privacy Policy
  • Sitemap