    Improving accuracy of cell and chromophore concentration measurements using optical density.
    BMC Biophys 2013 Apr 22;6(1). Epub 2013 Apr 22.
    Department of Chemical Engineering, Fenske Laboratory, The Pennsylvania State University, University Park 16802, PA.
    Background: UV-vis spectrophotometric optical density (OD) is the most commonly-used technique for estimating chromophore formation and cell concentration in liquid culture. OD wavelength is often chosen with little thought given to its effect on the quality of the measurement. Analysis of the contributions of absorption and scattering to the measured optical density provides a basis for understanding variability among spectrophotometers and enables a quantitative evaluation of the applicability of the Beer-Lambert law. This provides a rational approach for improving the accuracy of OD measurements used as a proxy for direct dry weight (DW), cell count, and pigment levels.

    Results: For pigmented organisms, the choice of OD wavelength presents a tradeoff between the robustness and the sensitivity of the measurement. The OD at a robust wavelength is primarily the result of light scattering and does not vary with culture conditions; whereas, the OD at a sensitive wavelength is additionally dependent on light absorption by the organism's pigments. Suitably robust and sensitive wavelengths are identified for a wide range of organisms by comparing their spectra to the true absorption spectra of dyes. The relative scattering contribution can be reduced either by measurement at higher OD, or by the addition of bovine serum albumin. Reduction of scattering or correlation with off-peak light attenuation provides for more accurate assessment of chromophore levels within cells. Conversion factors between DW, OD, and colony-forming unit density are tabulated for 17 diverse organisms to illustrate the scope of variability of these correlations. Finally, an inexpensive short pathlength LED-based flow cell is demonstrated for the online monitoring of growth in a bioreactor at culture concentrations greater than 5 grams dry weight per liter which would otherwise require off-line dilutions to obtain non-saturated OD measurements.

    Conclusions: OD is most accurate as a time-saving proxy measurement for biomass concentration when light attenuation is dominated by scattering. However, the applicability of OD-based correlations is highly dependent on the measurement specifications (spectrophotometer model and wavelength) and culture conditions (media type; growth stage; culture stress; cell/colony geometry; presence and concentration of secreted compounds). These variations highlight the importance of treating literature conversion factors as rough approximations as opposed to concrete constants. There is an opportunity to optimize measurements of cell pigment levels by considering scattering and absorption-dependent wavelengths of the OD spectrum.

    Similar Publications

    Optimal densitometry wavelengths that maximize radiochromic film sensitivity while minimizing OD growth and temperature sensitivity artifacts.
    J Xray Sci Technol 2009 ;17(1):61-73
    Department of Radiation Oncology, Oklahoma University Health Sciences Center, Oklahoma City, OK 73104, USA.
    It is well known that optical density (OD) of the radiochromic film (RCF) continues to grow after exposure at rates that have a complex dependence on dose, temperature, and densitometry wavelength. Dose rate and fractionation artifacts associated with variations in OD growth may limit the accuracy achievable by RCF dosimetry in brachytherapy and external beam applications, particularly at low doses (<5 Gy) and low dose rates (<10 cGy/h) where OD growth and sensitivity effects are large. To identify densitometry wavelengths that minimize OD growth artifacts and enhance RCF sensitivity at low doses, we have investigated Model MD-55-2 RCF response as a function of densitometry wavelength, irradiation-to-densitometry time interval, dose and temperature. Read More
    Robust metamodel-based inverse estimation of bulk optical properties of turbid media from spatially resolved diffuse reflectance measurements.
    Opt Express 2015 Oct;23(21):27880-98
    Estimation of the bulk optical properties of turbid samples from spatially resolved reflectance measurements remains challenging, as the relation between the bulk optical properties and the acquired spatially resolved reflectance profiles is influenced by wavelength-dependent properties of the measurement system. The resulting measurement noise is apparent in the estimation of the bulk optical properties. In this study, a constrained inverse metamodeling approach is proposed to overcome these problems. Read More
    The optimal wavelength of spectrophotometric broth microdilution antifungal susceptibility test for flucytosine and three azoles.
    Korean J Lab Med 2009 Aug;29(4):324-30
    Department of Laboratory Medicine, Inje University College of Medicine, Busan, Korea.
    Background: There is no guideline for the appropriate wavelength at which to measure the optical density (OD) value in broth microdilution antifungal susceptibility testing, although a spectrophotometric reading method is commonly used. The present study aimed to analyze the difference in the OD values over the range of visible light and to ascertain the optimal wavelength for the spectrophotometric method of microdilution testing.

    Methods: We measured the OD of background blank controls of broth medium, antifungal agents, and inocula of five type strains using a Synergy HT multi-detection microplate reader at 5-nm intervals from 380 nm to 760 nm. Read More
    Evaluation of optical coherence quantitation of analytes in turbid media by use of two wavelengths.
    Appl Opt 1999 Apr;38(10):2097-104
    Lawrence Livermore National Laboratory, 7000 East Avenue, Livermore, California 94550, USA.
    We introduce a novel method for determining analyte concentration as a function of depth in a highly scattering media by use of a dual-wavelength optical coherence tomography system. We account for the effect of scattering on the measured attenuation by using a second wavelength that is not absorbed by the sample. We assess the applicability of this technique by measuring the concentration of water in an Intralipid phantom, using a probe wavelength of 1. Read More