Prospective motion correction using inductively coupled wireless RF coils.

Magn Reson Med 2013 Sep 27;70(3):639-47. Epub 2013 Jun 27.

Department of Radiology, Stanford University, Stanford, California, USA.

Purpose: A novel prospective motion correction technique for brain MRI is presented that uses miniature wireless radio-frequency coils, or "wireless markers," for position tracking.

Methods: Each marker is free of traditional cable connections to the scanner. Instead, its signal is wirelessly linked to the MR receiver via inductive coupling with the head coil. Real-time tracking of rigid head motion is performed using a pair of glasses integrated with three wireless markers. A tracking pulse-sequence, combined with knowledge of the markers' unique geometrical arrangement, is used to measure their positions. Tracking data from the glasses is then used to prospectively update the orientation and position of the image-volume so that it follows the motion of the head.

Results: Wireless-marker position measurements were comparable to measurements using traditional wired radio-frequency tracking coils, with the standard deviation of the difference < 0.01 mm over the range of positions measured inside the head coil. Wireless-marker safety was verified with B1 maps and temperature measurements. Prospective motion correction was demonstrated in a 2D spin-echo scan while the subject performed a series of deliberate head rotations.

Conclusion: Prospective motion correction using wireless markers enables high quality images to be acquired even during bulk motions. Wireless markers are small, avoid radio-frequency safety risks from electrical cables, are not hampered by mechanical connections to the scanner, and require minimal setup times. These advantages may help to facilitate adoption in the clinic.

Download full-text PDF

Source
http://dx.doi.org/10.1002/mrm.24845DOI Listing
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC4006309PMC
September 2013
21 Reads

Publication Analysis

Top Keywords

motion correction
16
prospective motion
16
wireless markers
12
connections scanner
8
head coil
8
motion
5
wireless
5
safety risks
4
pulse-sequence combined
4
markers tracking
4
tracking pulse-sequence
4
enables high
4
range positions
4
knowledge markers'
4
unique geometrical
4
001 range
4
small avoid
4
markers' unique
4
radio-frequency safety
4
avoid radio-frequency
4

References

(Supplied by CrossRef)

Ackerman et al.
1986
Dynamic scan-plane tracking using MR position monitoring
Derbyshire et al.
J Magn Reson Imaging 1998

Krueger et al.
2006

Ooi et al.
2012

Aksoy et al.
2012

Beall et al.
2012

Haeberlin et al.
2012

Similar Publications