Transcatheter closure of patent ductus arteriosus: initial study on echocardiographic estimation of device size.

J Tehran Heart Cent 2010 30;5(4):199-201. Epub 2010 Sep 30.

Tehran Heart Center, Tehran University of Medical Sciences, Tehran, Iran.

Background: Transcatheter occlusion of the patent ductus arteriosus (PDA) is a minimally invasive treatment. The appropriate device size is chosen based on the angiographic measurement of the PDA. The current study aimed to assess the relationship between the transthoracic echocardiographic (TTE) measurements of the PDA prior to the occlusion procedure and the actual size of the deployed device.

Methods: We reviewed the available records of 7 patients (2 male) who underwent the procedure at our institution (mean age: 21 ± 12.7 years, range: 7 to 46 years). PDA closure was performed successfully using the Amplatzer Duct Occluder (n = 5) and its Chinese copycat, Cardi-O-Fix Occluder (n = 2).

Results: The TTE measurement of the aortic end diameter of the PDA showed a good linear regression correlation with the size of the implanted duct occluder [duct occluder size = 0.543 + (0.941× TTE measured diameter), R = 0.907; p value ≤ 0.01].

Conclusion: TTE can provide a good estimation of the size of the Amplatzer duct occluder.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC3466856PMC
October 2012

Publication Analysis

Top Keywords

duct occluder
12
patent ductus
8
ductus arteriosus
8
amplatzer duct
8
device size
8
size
6
pda
5
occluder
5
years pda
4
range years
4
years range
4
cardi-o-fix occluder
4
pda closure
4
closure performed
4
chinese copycat
4
copycat cardi-o-fix
4
performed amplatzer
4
occluder chinese
4
age 127
4
size deployed
4

Similar Publications