BD rapid serum tubes reduce false positive plasma troponin T results on the Roche Cobas e411 analyzer.

Clin Biochem 2012 Jul 26;45(10-11):842-4. Epub 2012 Apr 26.

Department of Laboratory Medicine and Pathology, Mayo Clinic, Rochester MN 55905, USA.

Objectives: In an attempt to reduce false positive results and improve turnaround time, we investigated the BD Rapid Serum Tube as an alternate sample type to lithium heparin plasma for Roche Troponin T analysis on the Roche Cobas e411 analyzer.

Design And Methods: BD Plasma Separator Tubes (PST) and Rapid Serum Tubes (RST) were collected in tandem from Emergency Department patients who had clinical orders for Troponin T over a 1 month period.

Results: RST and PST samples yielded analytically and clinically concordant Troponin T results on the Roche Cobas e411. Rare false positive results in lithium heparin samples were not observed with rapid clot serum tubes.

Conclusions: RST samples are appropriate for stat Troponin T analysis, and appear to reduce the incidence of rare false positive Troponin T results obtained with lithium heparin samples.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.clinbiochem.2012.04.014DOI Listing
July 2012
65 Reads

Publication Analysis

Top Keywords

false positive
16
roche cobas
12
lithium heparin
12
cobas e411
12
rapid serum
12
troponin analysis
8
troponin roche
8
heparin samples
8
rare false
8
serum tubes
8
reduce false
8
troponin
6
e411 analyzerdesign
4
type lithium
4
rapid clot
4
methods plasma
4
plasma separator
4
observed rapid
4
analyzerdesign methods
4
analytically clinically
4

Similar Publications