Dosimetric characterization of an image-guided stereotactic small animal irradiator.

Authors:
Strahinja Stojadinovic, Ph.D.
Strahinja Stojadinovic, Ph.D.
UT Southwestern Medical Center
Associate Professor
Medical Physics
Dallas, TX | United States

Phys Med Biol 2011 Apr 28;56(8):2585-99. Epub 2011 Mar 28.

Department of Radiation Oncology, University of Texas Southwestern Medical Center, Dallas, TX, USA.

Small animal irradiation provides an important tool used by preclinical studies to assess and optimize new treatment strategies such as stereotactic ablative radiotherapy. Characterization of radiation beams that are clinically and geometrically scaled for the small animal model is uniquely challenging for orthovoltage energies and minute field sizes. The irradiator employs a commercial x-ray device (XRAD 320, Precision x-ray, Inc.) with a custom collimation system to produce 1-10 mm diameter beams and a 50 mm reference beam. Absolute calibrations were performed using the AAPM TG-61 methodology. Beam's half-value layer (HVL) and timer error were measured with an ionization chamber. Percent depth dose (PDD), output factors (OFs) and off-axis ratios were measured using radiochromic film, a diode and a pinpoint ionization chamber at 19.76 and 24.76 cm source-to-surface distance (SSD). PDD measurements were also compared with Monte Carlo (MC) simulations. In-air and in-water absolute calibrations for the reference 50 mm diameter collimator at 19.76 cm SSD were measured as 20.96 and 20.79 Gy min(-1), respectively, agreeing within 0.8%. The HVL at 250 kVp and 15 mAs was measured to be 0.45 mm Cu. The reference field PDD MC simulation results agree with measured data within 3.5%. PDD data demonstrate typical increased penetration with increasing field size and SSD. For collimators larger than 5 mm in diameter, OFs measured using film, an ion chamber and a diode were within 3% agreement.

Download full-text PDF

Source
http://dx.doi.org/10.1088/0031-9155/56/8/016DOI Listing
April 2011
15 Reads
15 Citations
2.761 Impact Factor

Publication Analysis

Top Keywords

small animal
12
absolute calibrations
8
ionization chamber
8
measured
6
reference diameter
4
reference beam
4
beams reference
4
diameter collimator
4
1-10 diameter
4
diameter beams
4
calibrations reference
4
beam absolute
4
calibrations performed
4
simulations in-air
4
in-air in-water
4
carlo simulations
4
aapm tg-61
4
in-water absolute
4
performed aapm
4
collimator 1976
4

References

(Supplied by CrossRef)

Butson M J et al.
Phys. Med. Biol. 1998

Deng H et al.
Phys. Med. Biol. 2007

Similar Publications