An x-ray image guidance system for small animal stereotactic irradiation.

Authors:
Strahinja Stojadinovic, Ph.D.
Strahinja Stojadinovic, Ph.D.
UT Southwestern Medical Center
Associate Professor
Medical Physics
Dallas, TX | United States

Phys Med Biol 2010 Dec 16;55(23):7345-62. Epub 2010 Nov 16.

Department of Radiation Oncology, UT Southwestern Medical Center at Dallas, Dallas, TX 75235, USA.

An x-ray image-guided small animal stereotactic irradiator was developed and characterized to enable tumor visualization and accurate target localization for small field, high dose irradiation. The system utilizes a custom collimation system, a motorized positioning system (x, y, θ), a digital imaging panel and operating software, and is integrated with a commercial x-ray unit. The essential characteristics of the irradiator include small radiation fields (1-10 mm), high dose rate (>10 Gy min(-1)) and submillimeter target localization. The software enables computer-controlled image acquisition, stage motion and target localization providing simple and precise automated target localization. The imaging panel was characterized in terms of signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR) and spatial resolution. Overall localization accuracy and precision were assessed. SNR, CNR and spatial resolution are 24 dB, 21 dB and 2.8 lp mm(-1), respectively, and localization accuracy is approximately 65 µm with 6 µm precision. With the aid of image guidance, system performance was subsequently used to evaluate radiation response in a rat orthotopic lung tumor effectively sparing normal tissues and in a mouse normal lung. The capabilities of 3D treatment and cone-beam computed tomography are presented for 3D localization and delivery as a work in progress.

Download full-text PDF

Source
http://dx.doi.org/10.1088/0031-9155/55/23/011DOI Listing
December 2010
18 Reads
7 Citations
2.761 Impact Factor

Publication Analysis

Top Keywords

target localization
16
spatial resolution
8
small animal
8
animal stereotactic
8
imaging panel
8
cnr spatial
8
guidance system
8
localization accuracy
8
image guidance
8
high dose
8
localization
7
system
5
characterized terms
4
panel characterized
4
localization imaging
4
terms signal-to-noise
4
automated target
4
precise automated
4
contrast-to-noise ratio
4
ratio cnr
4

References

(Supplied by CrossRef)

De Salles A A et al.
AJNR Am. J. Neuroradiol. 1996

Deng H et al.
Phys. Med. Biol. 2007

Desrosiers C et al.
Technol. Cancer Res. Treat. 2003

Herfarth K K et al.
J. Clin. Oncol. 2001

Similar Publications