Sagittal curvature of total knee replacements predicts in vivo kinematics.

Authors:
Oliver Kessler
Oliver Kessler
University of Zurich
Switzerland
Scott Banks
Scott Banks
University of Florida
Henrich Mannel
Henrich Mannel
Biomet Deutschland GmbH
Frederic Marin
Frederic Marin
University of Ulm
Germany

Clin Biomech (Bristol, Avon) 2007 Jan 9;22(1):52-8. Epub 2006 Oct 9.

Scientific Affairs, Stryker Europe, Thalwil, Switzerland.

Background: It is known that in vivo kinematics after total knee replacement is influenced by the design of the implant. The goal of this study was to show that the sagittal curvature of two different knee prostheses differing in geometric design predicts their in vivo motion behavior.

Methods: Three-dimensional tibio-femoral displacements of two prosthesis designs (single radius vs. dual radius) were measured during knee extension under weight bearing conditions by in vivo video fluoroscopy. Finite helical axes were computed to represent the tibio-femoral motions. Angular deviation alpha and the spatial localization deviation delta were used to characterize the motions. Angular deviation is the angle between each incremental finite helical axis and the medio-lateral axis of the femoral component of the prosthesis. The spatial localization deviation is the distance between each finite helical axis and the center of the femoral component of the prosthesis. Statistical comparisons were performed using the median and the interquartile range of the angular deviation and the spatial localization deviation.

Findings: The single-radius design showed finite helical axes concentrated at a single axis near to the medio-lateral axis of the femoral component. The angular and spatial localization deviation of the dual radius design were larger compared to the single radius design, exhibiting finite helical axes varying between two axes.

Interpretation: Video fluoroscopy in combination with finite helical axis analysis proved to be suitable methods to evaluate the in vivo kinematical behavior of total knee arthroplasty, which can be useful for implant designers. Knowledge of in vivo kinematics can also provide surgeons with more background information about the total knee arthroplasty models they implant.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.clinbiomech.2006.07.011DOI Listing
January 2007
12 Reads

Publication Analysis

Top Keywords

finite helical
24
total knee
16
spatial localization
16
localization deviation
12
vivo kinematics
12
femoral component
12
helical axes
12
angular deviation
12
helical axis
12
single radius
8
medio-lateral axis
8
dual radius
8
motions angular
8
component prosthesis
8
predicts vivo
8
video fluoroscopy
8
sagittal curvature
8
knee arthroplasty
8
axis medio-lateral
8
axis femoral
8

Altmetric Statistics

Similar Publications

Contribution of geometric design parameters to knee implant performance: Conflicting impact of conformity on kinematics and contact mechanics.

Knee 2015 Jun 17;22(3):217-24. Epub 2015 Mar 17.

State Key Laboratory for Manufacturing System Engineering, School of Mechanical Engineering, Xi'an Jiaotong University, Xi'an, China; Institute of Medical and Biological Engineering, School of Mechanical Engineering, University of Leeds, Leeds, UK.

Background: Articular geometry of knee implant has a competing impact on kinematics and contact mechanics of total knee arthroplasty (TKA) such that geometry with lower contact pressure will impose more constraints on knee kinematics. The geometric parameters that may cause this competing effect have not been well understood. This study aimed to quantify the underlying relationships between implant geometry as input and its performance metrics as output. Read More

View Article
June 2015

In-vivo analysis of flexion axes of the knee: Femoral condylar motion during dynamic knee flexion.

Clin Biomech (Bristol, Avon) 2016 Feb 24;32:102-7. Epub 2015 Dec 24.

Bioengineering Laboratory, Department of Orthopaedic Surgery, Massachusetts General Hospital/Harvard Medical School, 55 Fruit Street, GRJ 1215, Boston, MA 02114, United States. Electronic address:

Background: Transepicondylar axis and geometrical center axis are widely used for investigation of the knee kinematics and component alignment in total knee arthroplasty. However, the kinematic characteristics of these knee axes are not well defined in literature. This study investigated the femoral condylar motion during a dynamic flexion of the knee using different flexion axes. Read More

View Article
February 2016

In vivo contact kinematics and contact forces of the knee after total knee arthroplasty during dynamic weight-bearing activities.

J Biomech 2008 Jul 5;41(10):2159-68. Epub 2008 Jun 5.

Bioengineering Laboratory, Orthopaedic Surgery, Massachusetts General Hospital/Harvard Medical School, Boston, MA 02114, USA.

Analysis of polyethylene component wear and implant loosening in total knee arthroplasty (TKA) requires precise knowledge of in vivo articular motion and loading conditions. This study presents a simultaneous in vivo measurement of tibiofemoral articular contact forces and contact kinematics in three TKA patients. These measurements were accomplished via a dual fluoroscopic imaging system and instrumented tibial implants, during dynamic single leg lunge and chair rising-sitting. Read More

View Article
July 2008

In vivo kinematics of knee replacement during daily living activities: Condylar and post-cam contact assessment by three-dimensional fluoroscopy and finite element analyses.

J Orthop Res 2017 07 19;35(7):1396-1403. Epub 2016 Sep 19.

Department of BEAMS, Université Libre de Bruxelles, Bruxelles, Belgium.

In total knee replacement, the investigation on the exact contact patterns at the post-cam in implanted patients from real in vivo data during daily living activities is fundamental for validating implant design concepts and assessing relevant performances. This study is aimed at verifying the restoration of natural tibio-femoral condylar kinematics by investigating the post-cam engagement at different motor tasks. An innovative validated technique, combining three-dimensional fluoroscopic and finite element analyses, was applied to measure joint kinematics during daily living activities in 15 patients implanted with guided motion posterior-stabilized total knee replacement. Read More

View Article
July 2017