Sample volume misregistration in linear array-based dual beam Doppler ultrasound systems.

IEEE Trans Ultrason Ferroelectr Freq Control 2003 Jul;50(7):836-47

School of Informatics, University of Wales, Bangor, Bangor LL57 1UT, UK.

Large velocity estimation errors can occur in dual beam Doppler ultrasound velocity measurement systems when there is left/right sample volume misregistration, particularly when the interbeam angle is small. Such misregistration will occur when there is tissue inhomogeneity. This is investigated for a typical type of inhomogeneity--a layer of fat--by calculating the amount of both angle and translation misregistration occurring in such a system realized using a single linear array transducer. The complex sample volume sensitivity is calculated using a modified time domain approach, combining the spatial impulse response method with ray tracing. The effects on these misregistrations of altering the aperture sizes and their relative positions on the array is then investigated to derive an improved aperture configuration for dual beam velocity estimation. Arrangements with transmit apertures wider than the receive apertures are shown to be preferable in this context.

Download full-text PDF

Source
http://dx.doi.org/10.1109/tuffc.2003.1214503DOI Listing
July 2003
3 Reads

Publication Analysis

Top Keywords

dual beam
12
sample volume
12
beam doppler
8
velocity estimation
8
doppler ultrasound
8
volume misregistration
8
calculated modified
4
inhomogeneity--a layer
4
sensitivity calculated
4
complex sample
4
volume sensitivity
4
modified time
4
time domain
4
spatial impulse
4
impulse response
4
combining spatial
4
approach combining
4
domain approach
4
transducer complex
4
array transducer
4

Similar Publications